Modular filter media test rig for fractional efficiency, differential pressure, dust holding capacity and gravimetrical efficiency in suction mode. Version HF with temperature and humidity control
MFP 3000 system
- Simultaneous particle measurement in the raw gas and clean gas
- Particle size measurements from 0.2 – 40 µm
- Measurement of Cn max= 10^6 particles/cm3 without dilution
- Internationally comparable measurement results
- Widespread distribution of the measurement system
- High reproducibility of the testing method
- Easy use of different test aerosols, e.g. SAE Fine and Coarse, NaCl/KCl, DEHS
- Highest raw gas concentrations of up to > 1000 mg/m3 (ISO Fine) or > 5000 mg/m3 (ISO Coarse) with measurement of the fraction separation efficiency for burden tests
- Flexible filter test software FTControl
- Sequence programs for pressure loss measurements, measurements of fraction separation efficiency and burden measurements
- Easy to operate, even untrained personnel can be instructed quickly in the use of the equipment
- MFP 3000 filter media inspection system with short installation time
- Cleaning and calibration can be performed autonomously by the customer
- Easy use of the measurement technology components – even in other applications
- Mobile setup, easy to move on castors
*Learn more about Palas: Here
*Fanpage: DIGI TECHNOLOGIES
Measurement range (size) | 0.2 – 40 µm |
---|---|
Measurement range (mass) | Up to 1,000 mg/m³ (depending on the version) |
Volume flow | 1 – 35 m3/h - suction mode |
Inflow velocity | 5 – 100 cm/s (others on request) |
Differential pressure measurement | 0 – 1,200 Pa selectable, 0 – 2,500 Pa selectable, 0 – 5,000 Pa selectable |
Aerosols | Dusts (e.g., SAE dusts), salts (e.g., NaCl, KCl), liquid aerosols (e.g., DEHS) |
Compressed air supply | 6 – 8 bar |
Test area of the medium | 100 cm2 |
Dimensions | 2.500 • 680 • 1.550 mm (H • B • T) |
- For filter media and small filter elements
- Product development and supervision during production
- Test based on ISO 11155-1 / DIN 71460-1 (cabin air filter)
- Test based on ISO 5011 (engine front air filter)
- Test based on ISO 5011 (engine front air filter)
- Other standards in different versions
- Fully automatic measurement of fractional efficiency, pressure drop curve, dust holding capacity and gravity performance
- The results are internationally comparable due to the system's high distribution