Fast and economical quality control and development of flat media with dust, oil or salt starting at 120 nm. Versions MFP 1000 HEPA for EN 1822-3 and ISO 29463-3
- Particle size measurements from 120 nm
- Internationally comparable measurement results
- Widespread distribution of the measurement system
- High reproducibility of the testing method
- Easy use of different test aerosols, e.g. SAE Fine and Coarse, NaCl/KCl, DEHS
- Flexible filter test software FTControl
- Sequence programs for pressure loss measurements, measurements of fraction separation efficiency and burden measurements
- Easy to operate; even untrained personnel can be instructed quickly in the use of the equipment
- Short set-up times
- Cleaning and calibration can be performed autonomously by the customer
- Easy use of the measurement technology components – even in other applications
- Mobile setup, easy to move on castors
- Reliable operation
- Validation of the clear function of individual components and the overall system during pre-delivery acceptance testing and upon delivery
- Low-maintenance
Technical Data | Values |
Measurement range (size) | 0.12 – 40 µm |
Volume flow | 1 – 35 m3/h (pressurized operation) |
Power supply | 115/230 V; 50/60 Hz |
Dimensions | Approx. 600 • 1,800 • 900 mm (W • H • D) |
Inflow velocity | 5 – 100 cm/s (others on request) |
Differential pressure measurement | 0 – 2,500 Pa (others on request) |
Test area of the medium | 100 cm² |
Aerosol | Dusts (e. g. SAE dusts), salts (e. g. NaCl, KCl), liquid aerosols (e. g. DEHS) |
Aerosol concentrations | For SAE Fine without additional dilution up to 1,000 mg/m3 (ISO A2 Fine) |
Compressed air supply | 6 – 8 bar |
- For filter media and small mini-filters
- product development and during production monitoring
- Fast and cost efficient testing of fractional efficiency with dust, oil or salt based on ISO 11155-1 (cabin air filters), ISO 5011 (engine pre-air filters), EN 779/ Ashrae 52.2/ ISO 16890 (room air filters),
- Within the version MFP 100 HEPA measurement of MPPS of HEPA/ ULPA filter media based on EN 1822 and ISO 29463-3