Dilution system with dilution factor 1:100 The Palas® VKL 100 dilution systems are used in vertical operation for the particle size range up to 2 µm for applications in the clean room. Dilution factors of up to 1:100,000 are achieved by cascading several VKL systems.
- The dilution systems from Palas® are characterized unambiguously. This is documented with a calibration certificate for each individual device.
- The dilution steps deliver a temporally constant, representative dilution with the factors 10 and 100.
- The dilution systems can be cascaded with the factors 100, 1,000, 10,000 and 100,000
- Low compressed air consumption, e.g. just 128 l/minwith a dilution factor of 10,000 with four VKL 10 systems
- The dilution steps are combinable with all common particle counters.
- These cascaded dilution systems can be tested by the users themselves with a simple test set-up.
- Isobaric dilution up to 10 bar overpressure / isothermal dilution up to 120°C with the VKL 10 E, VKL 10 ED, KHG 10 and KHG 10 D dilution systems
- Simple functional test on-site
Technical Data |
Values |
Dimensions | 100 x 245 x 100 mm |
Khối lượng | Approx. 4 kg |
Dilution factor | 1:100 |
Isokinetic suction nozzles | 0.028 – 0.06 l/min,
0.23 – 0.5 l/min, 0.6 – 1.6 l/min, 2 – 5 l/min, 28 l/min => 15 – 37 l/min |
Maximum particle size | < 2 µm (for dusts) |
Special features | Cascadable |
Volume flow (clean air) | 17 – 45 l/min |
Volume flow (suction flow) | 0.15 – 0.5 l/min |
Compressed air supply | 4 – 8 bar |
- Aerosol measurement technology: test aerosols from filters and inertial separators
- Separation efficiency determination with counting measuring methods, e.g. HEPA/ULPA filters
- Leak test and acceptance measurements of clean rooms, isolators and safety work benches
- Inhalation toxicology
- Quality control of respirator masks and filter cartridges