Programmable DC Power Supplies Model 9151 | Electrical measurement of BK Precision
Product code:BK 9151
Supplier:BK Precision
Price:
Inquiry
The 9151 series are high performance linear-regulated programmable DC power supplies that provides excellent performance and features not found in other supplies in this price category. These supplies are designed for applications in design verification, production or university labs that require high yet clean and reliable power, combined with excellent resolution, accuracy and fast transient response time.
- Very low ripple and noise due to linear regulation
- Excellent display resolution
- Fast transient response time (<120 µs)
- SCPI compatible
- Communicate via USB Interface, using included USB to TTL serial converter cable
- List mode operation for increased throughput
- Very quiet due to intelligent fan speed control
- For bench use or rack mountable
- Application Software for front panel emulation and simple test sequence generation included
- Digital port with Remote Inhibit (RI) useful for turning multiple power supplies On/Off simultaneously
- Over voltage protection
- Digital Volt Meter and Milliohm Meter
MODELS | 9150 | 9151 | 9152 | 9153 |
Output Voltage | 0 - 5.2 V | 0 - 20 V | 0 - 30 V | 0 - 60 V |
Output Current | 0-60 A | 0-27 A | 0-18 A | 0-9 A |
Specifications |
9150 | 9151 | 9152 | 9153 |
Output ratings ( 0 °C ~ 40 °C) |
0 - 5.2 V 0 - 60 A |
0 - 20 V 0 - 27 A |
0 - 30 V 0 - 18 A |
0 - 60 V 0 - 9 A |
Load regulation ±(% of output+offset) |
< 0.01% + 2 mV < 0.1% + 10 mA |
< 0.01% + 1 mV < 0.1% + 5 mA |
< 0.01% + 1 mV < 0.1% + 2 mA |
|
Line regulation ±(% of output+offset) |
< 0.02% + 0.1 mV < 0.1% + 1 mA |
< 0.02% + 1 mV < 0.01% + 1 mA |
< 0.02% + 1 mV < 0.01% + 0.1 mA |
|
Programming resolution | 0.1 mV 1 mA |
1 mV 1 mA |
1 mV 0.1 mA |
|
Readback/meter resolution | 0.1 mV 1 mA |
0.1 mV 0.1 mA |
0.1 mV 0.1 mA |
|
Front panel setting resolution | 0.1 mV 1 mA |
0.5 mV 1 mA |
0.5 mV 1 mA |
|
Programming accuracy 12 months (25 °C ± 5 °C) ±(% of output+offset) |
< 0.02%+2 mV < 0.1%+30 mA |
< 0.02%+6 mV < 0.1%+15 mA |
< 0.02%+12 mV < 0.05%+10 mA |
|
Readback/ meter accuracy 12 months (25 °C ± 5 °C) ±(% of output+offset) |
< 0.02%+1.5 mV < 0.05%+15 mA |
< 0.02%+3 mV < 0.05%+10 mA |
< 0.02%+6 mV < 0.05%+5 mA |
|
Ripple & noise (20 Hz ~ 20 MHz) |
≤4 mVp-p 15 mArms |
≤4 mVp-p 5 mArms |
≤ 5 mVp-p 3 mArms |
|
Temperature coefficient, (0 °C~40 °C) ±(% of output+offset) |
< 0.02%+2 mV < 0.1%+30 mA |
< 0.02%+5 mV < 0.1%+15 mA |
< 0.02%+10 mV < 0.05%+5 mA |
|
Readback temperature coefficient, ±(% of output+offset) |
< 0.02%+2 mV < 0.1%+20 mA |
< 0.02%+5 mV ≤0.05%+10 mA |
< 0.02%+10 mV ≤0.05%+5 mA |
|
Transient response (for a change from 0 to 50% of maximum rated current) | 100 μs for output to recover to within 75 mV |
120 μs for output to recover to within 75 mV |
100 μs for output to recover to within 50 mV |
50 μs for output to recover to within 50 mV |
DVM accuracy | 0~12V range: 0.02%+2mV 0~40V range: 0.02%+3mV |
|||
DVM resolution | 0~12V range: 0.1mV 0~40V range: 1mV |
|||
Milliohm meter accuracy | 0.6% (for voltage and current ≥11% of full scale) 1.8% (for voltage and current ≥4% of full scale) |
|||
AC input | 115 V/220 VAC ± 10%, 47 to 63 Hz | |||
Weight | 63.9 lbs, (29 kg) | |||
Dimensions (W x H x D) | 16.88” x 3.47” x 18.06” (429 x 88 x 459 mm) | |||
Three-Year Warranty |
||||
Standard accessories | User manual, power cord, USB to TTL serial converter, application software installation disk, and certificate of calibration | |||
Optional accessories | IT-E151 rack mount kit |
The included Application Software supports front panel emulation and allows users to generate simple test sequences without the need to write source code.